Moisture


Weigh the crucible and record the value as M1,Take 2g of test samples in the crucible (correct to 0.0001g) And record the value as M2,put the testsamplesin the oven at 105℃ for 3h, cool in a dryer to room temperature, weigh and record the value as M3, put the testsamplesinto the muffle furnace and burn at 450℃ for 3h, cool in a dryer to room temperature,weigh and record the value as M4.

Cellulose Test Method

Instruments and equipments

➢ 30mL crucible
➢ oven
➢ muffle furnace
➢ electronic balance
➢ drier

Test procedures

Weigh the crucible and record the value as M1,Take 2g of test samples in the crucible (correct to 0.0001g) And record the value as M2,put the testsamplesin the oven at 105℃ for 3h, cool in a dryer to room temperature, weigh and record the value as M3, put the testsamplesinto the muffle furnace and burn at 450℃ for 3h, cool in a dryer to room temperature,weigh and record the value as M4.

Calculate

Moisture in mass fraction is calculated as follows.

W (Moisture) = [1-(M3-M1)/M2]*100

Make 2 parallel samples for each sample, average the results. If the difference between the two resultsis greater than 0.5, retest isrequired.

Ash


Weigh the crucible and record the value as M1,Take 2g of test samples in the crucible (correct to 0.0001g) And record the value as M2,put the testsamplesin the oven at 105℃ for 3h, cool in a dryer to room temperature, weigh and record the value as M3, put the testsamplesinto the muffle furnace and burn at 450℃ for 3h, cool in a dryer to room temperature,weigh and record the value as M4.

Ash

Instruments and equipments

➢ 30mL crucible
➢ oven
➢ muffle furnace
➢ electronic balance
➢ drier

Test procedures

Weigh the crucible and record the value as M1,Take 2g of test samples in the crucible (correct to 0.0001g) And record the value as M2,put the testsamplesin the oven at 105℃ for 3h, cool in a dryer to room temperature, weigh and record the value as M3, put the testsamplesinto the muffle furnace and burn at 450℃ for 3h, cool in a dryer to room temperature,weigh and record the value as M4.

Calculate

Ash in mass fraction is calculated as follows

W (Ash) = (M4-M1)/M2*100

Make 2 parallel samples for each sample, average the results. If the difference between the two resultsis greater than 0.5, retest isrequired.

Wet viscosity


Put the beaker on the scale and make zero, Take 2g of test samples in the beaker(correct to 0.01g) , add 198g of water at about 90 ° C. Stir quickly with a glass rod,make it dispersed uniformly,Then put the beaker in basin with ice water, Stir and move from fast to slow to preventsplashing. When the water cool to room temperature, take out the beaker and dry the surface, putit onthe scale thathasbeenzeroed, according to the actual mass of the sample

Wet viscosity

Instruments and equipments

➢ Brookfield type rotary viscometer
➢ 250mL High beaker
➢ Thermostat water bath
➢ Thermometer
➢ Ice
➢ Basin
➢ PH meter or PH test strips

Test procedures

Put the beaker on the scale and make zero, Take 2g of test samples in the beaker(correct to 0.01g) , add 198g of water at about 90 ° C. Stir quickly with a glass rod,make it dispersed uniformly,Then put the beaker in basin with ice water, Stir and move from fast to slow to preventsplashing. When the water cool to room temperature, take out the beaker and dry the surface, putit onthe scale thathasbeenzeroed, according to the actual mass of the sample, add water to a 2% solution (High temperatures make water evaporate quickly, and will reduce the quality of water). Then put the beaker in basin with ice water for about 20 minutes, make it dissolved fully, then take the solution out of the beaker into thermostat water bath to 20℃±0.1℃, measure with the Brookfield viscometer (2% HPMC Solution,Spindle:No.7,rotate speed:20RPS).

Calculate

The viscosity of the solution is calculated as follows

η=k*α

Notes:

η——The viscosity of the solution, unit mPa.s ;

k——Rotor coefficient ;

α——Pointer value. Testtwiceofeachsampletogettwotestresults,iftheviscosity≤1000 mPa.s, parallelresults ofthe

relative deviation is not more than 5%;ifthe viscosity> 1000 mPa.s, parallelresults ofthe relative deviation is not more than 10%, the result is taken as the arithmetic average.

After testing the viscosity,test the PH of the sample solution with a PH meter or PH test strips and record.

Purity


Put the beaker on the scale and make zero, Take 2g of test samples in the beaker(correct to 0.01g) , add 198g of water at about 90 ° C. Stir quickly with a glass rod,make it dispersed uniformly,Then put the beaker in basin with ice water, Stir and move from fast to slow to preventsplashing. When the water cool to room temperature, take out the beaker and dry the surface, putit onthe scale thathasbeenzeroed, according to the actual mass of the sample

PH value

Instruments and equipments

➢ Brookfield type rotary viscometer
➢ 250mL High beaker
➢ Thermostat water bath
➢ Thermometer
➢ Ice
➢ Basin
➢ PH meter or PH test strips

Test procedures

Put the beaker on the scale and make zero, Take 2g of test samples in the beaker(correct to 0.01g) , add 198g of water at about 90 ° C. Stir quickly with a glass rod,make it dispersed uniformly,Then put the beaker in basin with ice water, Stir and move from fast to slow to preventsplashing. When the water cool to room temperature, take out the beaker and dry the surface, putit onthe scale thathasbeenzeroed, according to the actual mass of the sample, add water to a 2% solution (High temperatures make water evaporate quickly, and will reduce the quality of water). Then put the beaker in basin with ice water for about 20 minutes, make it dissolved fully, then take the solution out of the beaker into thermostat water bath to 20℃±0.1℃, measure with the Brookfield viscometer (2% HPMC Solution,Spindle:No.7,rotate speed:20RPS).

Calculate

The viscosity of the solution is calculated as follows

η=k*α

Notes:

η——The viscosity of the solution, unit mPa.s ;

k——Rotor coefficient ;

α——Pointer value. Testtwiceofeachsampletogettwotestresults,iftheviscosity≤1000 mPa.s, parallelresults ofthe

relative deviation is not more than 5%;ifthe viscosity> 1000 mPa.s, parallelresults ofthe relative deviation is not more than 10%, the result is taken as the arithmetic average.

After testing the viscosity,test the PH of the sample solution with a PH meter or PH test strips and record.